Table 1 Performance comparison of preprocessing methods.

From: LIMPIC: a computational method for the separation of protein MALDI-TOF-MS signals from noise

Protein concentration

2.5 fmol

5 fmol

7.5 fmol

10 fmol

25 fmol

50 fmol

LIMPIC

Average intensity

1.17

4.65

5.32

17.07

94.03

277.37

 

Standard deviation

0.10

0.39

0.61

2.70

14.01

42.48

 

Coefficient of variation

0.09

0.08

0.12

0.16

0.15

0.15

APEX & CENTROID

Average intensity

1.31

5.36

6.05

14.94

88.48

283.08

 

Standard deviation

0.32

0.74

0.97

2.40

12.76

42.48

 

Coefficient of variation

0.24

0.14

0.16

0.16>

0.14

0.15

CROMWELL

Average intensity

0

0

4.01

12.76

85.34

275.50

 

Standard deviation

0

0

1.52

2.95

13.79

48.49

 

Coefficient of variation

-

-

0.38

0.23

0.16

0.17

  1. LIMPIC, APEX, CENTROID and CROMWELL preprocessing methods were evaluated from in-vitro protein spectra at different dilutions of cytochrome C. For each method, the average peak intensity, the standard deviation and the coefficient of variation are provided. The underlined values were manually obtained from the mass spectra, because the peak-picking algorithms did not allow to detect the peak.